Analysis Method
Complete ZAF analysis is possible, with or without standards, using an internal database of fundamental parameters (FP) such as absorption coefficients, fluorescence yields, transition probabilities, etc. Analysis can be done without standards if the results can be normalized to 100%. When using standards, thickness can also be determined or the results do not have to be normalized.
When more than one excitation is used, at least one of the elements for each condition must have been calibrated. Calibration factors may be generated using any type of standard (e.g., pure element or analytical “type” standard). A single “type” standard may be used, or the calibration may be done with a different standard for each element, or any combination of standards may be used. If some elements are calibrated and some are not, the latter can use calibration coefficients derived from the former group.
The mass thickness of the sample can either be specified or calculated. If the latter, then the analysis cannot be standardless. Several units are possible for thickness measurement, and the density can be calculated theoretically or specified in the case of linear thickness calculations. Composition units may be ppm or wt%, with the additional output of atomic and mole percent.
Calibration Method
With the FP analysis method, one can choose a “standardless” approach. All of the parameters describing the X-ray tube spectra, filtering, attenuation in air, attenuation in Be windows and dead layers, attenuation and enhancement in the sample, etc. are computed from physical models based on the data the user has entered into the software. It is simple to use standardless analysis but the parameters are only approximate. This is due to approximations inherent in the physical models and in the data the user enters.
With the FP analysis method, one can also choose to calibrate its parameters using either a single standard or multiple standards. Calibration is strongly recommended and will lead to much more accurate analysis results. A single “type” standard may be used, i.e. one can use a single piece of material containing all of the elements which will later be analyzed. For example, one can use a single “standard reference material” of stainless steel and then obtain very accurate analyses of other steel alloys. One can also calibrate with a different standard for each element.
Several types of analysis cannot be standardless, i.e. calibration with standards (reference materials) is required. Least squares analysis cannot be standardless. If the mass thickness of the sample (i.e. the mg/cm2) is calculated, the analysis cannot be standardless.
Detectors
Various detectors (Si-PIN, SDD) and windows (Be, Si3N4) can be fully modeled. The software has provision for the user to input all the required parameters (e.g., thickness, are, dead layer, etc.) associated with these detectors and their windows.
Amptek supplies all the parameters for its XR-100 series of detectors.
Geometry
The complete system geometry can be specified including the sample incidence and take-off angles, the source-to-optic and/or source-to-sample distances, the sample-to-detector distance, as well as the environmental factors.
Figure 4. Geometry angle definitions.
Elements, Lines, and Interelement corrections
Includes full corrections for absorption and both thick and thin-film secondary fluorescence. All possible lines are considered for both excitation and fluorescence. The analysis can be performed for all elements from H through Fm, using K, L or M lines in the energy range from 0.1 keV up to 60 keV.