Software Overview
DPPMCA is a Windows software application that provides data acquisition, display, and control for Amptek signal processors. This software is free and can be downloaded from this link. Amptek highly recommends using the Professional version of any Microsoft operating system (Windows 7 Pro, Windows 10 Pro, etc.).
DPPMCA is compatible with FW6 devices only:
- DP5
- DP5G
- X-123 (with DP5 inside) including X-123SDD and X-123CdTe
- PX5
- PX5-HPGe
- Gamma-Rad5
- TB-5
- MCA-8000D including Option PA
For DPP installation instructions Follow this link.
Features
- Full control of all hardware features. Typically includes acquisition presets, gain, peaking time, detector HV, detector temperature, number of channels, risetime discrimination parameters, and many more.
- Live display of the spectrum with many options. Includes linear and logarithmic vertical scaling, manual or auto-ranging, and zoom on the horizontal scale. Displays live spectrum and multiple stored or processed spectra.
- Spectral analysis features include energy calibration, setting regions of interest (ROI), computing ROI information (centroid, total area, net area, FWHM), spectrum smoothing, summing of spectra, subtraction and scaling of background spectra.
- Active link to the XRS-FP Quantitative Analysis Software Package using fundamental parameters, which can be purchased separately.
- Supports Microsoft Windows XP, Vista, Windows 7 and Windows 10.
Hardware Control
All hardware parameters can be accessed by clicking the “Acquisition Setup” toolbar button. Controls are grouped by tabs. “MCA” tab controls acquisition presets, number of channels, LLD (threshold), setup of single channel analyzers (SCA’s), etc. “Shaping” tab controls peaking time, flat top, baseline restoration, pile-up rejection, and risetime discrimination. Common controls also set by shortcuts. Gain can be controlled by increment and decrement buttons on the toolbar, or by setting a specific value in the Setup display. “AutoTune” controls are available to automatically tune key parameters, such as the input offset and the fast and slow thresholds. Customized hardware configurations may be stored as .cfg files. DPPMCA is shipped with a library of .cfg files for the common configurations of Amptek’s detectors.
Display and User Interface
All functions can be accessed via the command menus. Many functions are also accessed by toolbar buttons. Includes “hot keys” for the most common functions, i.e. starting and stopping acquisition, rescaling the display, setting ROIs, etc. User control of display parameters such as log vs. lin scaling, auto-scaling, zoom along the energy axis, setting colors and grids, and many more. Oscilloscope mode displays pulse shapes for tuning of hardware parameters.
Spectrum Analysis
User can set regions of interest (ROIs) for peaks. Software computes centroids, FWHM, net area, and other quantities. Data are display on the screen for he selected ROI or in a table for all ROIs. Simple energy calibration, using two points or a linear regression for multiple points. Typically based on the centroids of known ROIs. Includes capability for simple arithmetic manipulation, e.g. subtracting a “blank” or “background” spectrum, scaling to match previous data, etc. For quantitative X-ray analysis, DPPMCA includes an active link to the XRF-FP software sold by Amptek. For other sophisticated spectral analyses, the data are saved in an ASCII format, that most software will recognize. Data can also be copied to the clipboard and pasted, e.g. in Excel.
File Formats
Data saved as .MCA file (in an ASCII format), which includes the raw spectrum data, a header with calibration and ROI information, and a footer containing the hardware configuration parameters. The ASCII data can be read by most standard analysis software packages.
System Requirements
Supports Microsoft Windows XP SP3, Vista, Windows 7, Windows 8, Windows 10. DPPMCA will run under both 32 and 64-bit versions.
Using DPPMCA
Installation
After downloading the .zip file, extract the executable "DppMCA_Setup_X.X.X.XX.exe" wherever is convenient and run it. Windows may prompt you with a warning which will require you to click "more info" then "run anyways". You will need to accept the End-User License Agreement and select an install directory.
For Digital Pulse Processors (DPP) like the DP5, PX5, X-123, or Gamma-Rad5, you will have to install the WINUSB driver. Please see the DPP installation instructions.
Quick Use
Please read the Quick Start Guide for your product. It contains specific instructions for the operation of your device. The information below is generic to DPPMCA. There are different options depending on the product in use.
Changing the Gain: This changes the full scale energy range.
A common task is to change the energy range of a system. This is done by adjusting the gain of the amplifier. A lower gain corresponds to a higher energy range, and a higher gain to a lower energy range. For example, in Amptek’s digital systems a gain of x100 may correspond to a 15 keV full scale and a gain of x50 to a 30 keV full scale. To adjust the energy range it is always the gain of the system that must be changed. It is not enough to change the software calibration only. The software calibration is only in software. If the software’s calibration is changed without the amplifier’s gain changing appropriately, the software will be displaying the wrong energy range and peaks will be incorrectly identified.
If you are using one of Amptek’s digital processors (DP5, X-123, Gamma-Rad5 etc.), the gain is adjusted through the DPPMCA software on the Gain & Pole Zero tab of the DPP Properties, or by using the gain buttons on the toolbar It is necessary to readjust the thresholds whenever the gain is changed. This can be done automatically by starting an acquisition (pressing space bar) and clicking the “Tune Slow/Fast Thresholds” button on the toolbar of the DPPMCA application with no source exciting the detector. See the quick start guide of the appropriate product for more information.
Calibration: Changing the Channel scale to Energy.
Once the appropriate gain (full scale energy) is found, the software can be calibrated. The software must always be calibrated in order to change the channel scale into an energy scale. To accurately calibrate a spectrum there must be at least two known peaks in the spectrum. For example, an iron (Fe) peak at 6.40 keV and molybdenum (Mo) peak at 17.48 keV. It is best to use peaks that are on the two ends of the energy range of interest. Peaks that are too close together may not produce an accurate calibration. For example, do not use Mn at 5.9 keV and Fe at 6.4 keV to calibrate a 30 keV full scale.
To calibrate the scale in the Amptek DPPMCA software follow these steps:
- Chose the two peaks that will be used for the calibration and make note of their energies.
- Mark an ROI (region of interest) around each peak. This can be done by clicking the cursor at the left base of the first peak and then holding down the “U” key on the keyboard until the whole peak is highlighted, or by using the Edit ROI button on the toolbar and then clicking and dragging across the peak until it is highlighted. For more information on marking ROIs please see the Help file under Analyzing Data with the DPPMCA Software -> Regions of Interest and How to Define Them.
- Now click in the toolbar button “Calibrate” to open the calibration dialog box.
- Move the dialog box such that both peaks are visible and then click the cursor into the first peak. The peak should be highlighted and the Peak Information section on the right-hand information panel should be filled in. If the peak is highlighted and there is no peak information, then the ROI is not marked correctly. Try adjusting the start and end points of the ROI.
- Click the Centroid button on the dialog box. This will enter the center position of the peak into the Channel box. Then enter the energy of that peak into the Value box, e.g. 6.4, and click Add.
- Now click the cursor into the second peak (it will highlight). Click on Centroid, enter the second energy, e.g. 17.48, and click Add. There should now be two lines in the box.
- In the Units box type in the energy units, e.g. keV.
- Click OK. The scale should now be in energy. If it is not press F7, this toggles between calibrated units and channels.